Enhance your patient safety and simplify your workflow. The HemoCue® Hb 201 DM System offers the lab accuracy and ease of use that has become a standard in Hb point-of-care testing, while also providing improved patient safety with features like QC lockout and built-in barcode scanning of patient ID.
Key benefits
Improve your administrative workflow and safeguard patient testing and data with our data management solution in hemoglobin point-of-care testing. The HemoCue® Hb 201 DM System offers a complete data management solution for Hb testing and anemia screening through bi-directional communication and flexible settings.
01
Results within 1 minute
02
Built-in quality control self-test
03
Link patient information and comments with test results
04
Reduce errors with automatic result transfer
05
Seamlessly interface with existing network using POCT1-A
06
Easy to use after brief training
* System specifications :
- Principle : Modified azidemethemoglobin reaction; dual wavelengths (570 nm and 880 nm) for compensation of turbidity .
- Calibration : Calibrated against the hemiglobincyanide (HiCN) method, the international reference method recommended by the ICSH, for the determination of the hemoglobin concentration in blood. The system is factory calibrated and needs no further calibration .
- Power : Internal rechargeable batteries or powered with AC adapter direct or in docking station
- Quality control : Built-in self-test, optional liquid controls
- Sample material : Capillary, venous or arterial whole blood
- Sample volume : ~10 µL
- Measurement range : 0.5-25.6 g/dl (5-256 g/L, 0.3-15.9 mmol/L)
- Measuring time : 15-60 seconds
Dimensions
Analyzer:170 x 93 x 50 mm (6.70 x 3.66 x 1.97 inches)
Docking station:135 x 206 x 61 mm (5.30 x 8.10 x 2.40 inches)
Weight
Analyzer:350 g (0.77 lbs) with batteries installed
Docking station:566 g (1.24 pounds)
Storage temperature
Analyzer:0-50°C (32-122°F)
Microcuvettes:unopened 15-30°C (59-86°F); three-month open vial stability
Operating temperature
18-30°C (64-86°F)
Data management features
Data entry
Touchscreen display, built-in barcode scanner*
Interface capabilities
Interface with existing software solutions using POCT1-A (CLSI standard). Docking solution enables connection of up to 5 analyzers
Configurable functions
Operator ID, Patient ID, Lot ID, patient test comments, critical value alerts, STAT testing
Results storage
4.000 patient/STAT tests, 500 QC tests, 500 analyzer logs, 200 patient IDs
Quality control
QC lock-out
Data management
Linearity and proficiency testing