HemoCue®
Glucose 201 RT System
The HemoCue® Glucose 201 RT System enables you to make the right decisions regarding your patient – at the point of care. Within only a minute, you have lab-quality answers to help put your patient at ease. With the added benefit of room temperature microcuvette storage, the HemoCue® Glucose 201 RT System simplifies your work process.
*Not available in the US, for corresponding US product, see hemocue.us
Key benefits
The HemoCue® Glucose 201 RT System is a portable and battery-operated system that helps you reduce margins of error and risk of infection, offering convenience and flexibility.
01
Results within 1 minute
02
Precise monitoring for better glycemic control
03
Microcuvette technology means reducing the risk of infection spreading
04
Individually wrapped microcuvettes means minimizing contamination
05
Factory calibrated analyzers mean no need to recalibrate
06
Portable and battery-operated system with room temperature microcuvette storage
* System specifications :
- Principle : Modified glucose degydrogenase in which the total amount of glucose is measured photometrically at the end point
- Calibration : Factory calibrated and traceable to the ID GC-MS method; needs no further calibration and no coding
- Sample material : Capillary, venous or arterial whole blood
- Sample volume : <4 µL
- Measurement range : Plasma equivalent values: 0.61-30.8 mmol/L (11-555 mg/dL)
- Whole blood values: 0.55-27.8 mmol/L (10-500 mg/dL)
- Measuring time : Within 1 minute for normal glucose levels
- Dimensions : 160 x 85 x 43 mm (6.30 x 3.35 x 1.69 inches)
- Weight : 350 g (0.77 pounds) with batteries installed
- Storage temperature : Analyzer: 0-50°C (32-122°F)
- Microcuvettes: 0-30°C (32-86°F)
- Operating temperature : 15-27°C (59-80°F)
- Power : AC Adapter or 4 AA batteries
- Interface : Printer and HemoCue® Basic connect including optional barcode scanner. Data transfer using Bluetooth® technology is possible via HemoCue® BT Connect (accessory)
- Quality control : Built-in self-test, system can be verified using liquid controls