HemoCue®
Glucose 201+ System
The HemoCue® Glucose 201+ System puts lab-quality answers in your hands when they’re needed most – at the point of care. Not only does the unique microcuvette technology enable accurate result reading in only three simple steps, it also reduces contamination risk.
*Not available in the US, for corresponding US product, see hemocue.us
Key benefits
Used in screening, monitoring and as an aid in the diagnosis of diabetes mellitus, the HemoCue® Glucose 201+ System enables you to make the right decision at the point of care while reducing margins of error and risk of infection.
01
Results within 1 minute
02
Precise monitoring for better glycemic control
03
Possible to detect hypoglycemia in newborns
04
Factory calibrated analyzers means no need to recalibrate
05
Portable and battery-operated system means ideal for mobile settings
06
Microcuvette technology means reduced risk of infection spreading
* System specifications :
-Principle : Modified glucose degydrogenase in which the total amount of glucose is measured at the end point photometrically
- Calibration : Factory calibrated and traceable to the ID GC-MS method; needs no further calibration and no coding
- Power : AC Adapter or 4 AA batteries
- Interface : Printer and HemoCue® Basic Connect including optional barcode scanner. Data transfer using Bluetooth® technology is possible via HemoCue® BT Connect (accessory)
- Measurement range : Plasma equivalent values: 0.61-24.6 mmol/L (11-444 mg/dL)
- Whole blood values : 0.55-22.2 mmol/L (10-400 mg/dL)
- Measuring time : Within 1 minute for normal glucose levels
- Sample material : Capillary, venous or arterial whole blood
- Sample volume : < 5 µL
- Operating temperature : 15-30°C (59-86°F)
* Storage temperature :
- Analyzer : 0-50°C (32-122°F)
- Microcuvettes: unopened below 8°C (46°F) (incl. storage in a freezer), room temperature for up to 3 days; one-month open vial stability
- Dimensions
160 x 85 x 43 mm
(6.30 x 3.35 x 1.69 inches)
- Weight : 350 g (0.77 pounds) with batteries installed
- Quality control : Built-in self-test, system can be verified using liquid controls