-Professional monitor/defibrillators:
An advanced, portable, multi-parameter vital signs monitor with integrated telemedicine. Tempus IC2 allows trained non-medical users to manage medical emergencies in the air, at sea, and in remote locations. Along with built-in ease of use, the Tempus IC2 features clear visual guidance and automated data collection, helping users to feel confident in managing a medical emergency.
-Real-time access to support:
Tempus IC2 allows intermittent users and non-medical professionals with basic medical training to obtain and transmit a patient's clinical vital signs data, including 12-lead ECG plus images and video. The device
also offers a two-way voice connection to connect the user to professional medical support who can access a patient remotely.
-Discreet access to medical infrastructure:
No one can predict when or where they may be injured or taken ill, or how serious the situation may be. Tempus IC2 provides assistance to medical professionals through a reliable, secure communication link with voice and data automatically connecting via existing satellite or terrestrial communication systems, with no need for proprietary software or standalone equipment.
- Credible information, clear processes:
With its compact, lightweight and robust design, Tempus IC2 can be used in any medical emergency situation. The collection of clinical grade, vital signs information that is routinely captured in an emergency room along with voice connection, visual guidance and logical help screens allows users wherever they are in the world to obtain professional support with difficult decisions, treatment guidance and advice.
-Connected medical support:
Imagine being in a remote and challenging location and hours away from quality medical help. With Tempus IC2, users can securely connect 24/7 to medical professionals who can fully control Tempus IC2 remotely while seeing exactly the same information on their screens as those displayed on the Tempus IC2, providing much-needed peace of mind to the user and patient.
Specifications*
- Standalone size: 263mm x 216mm x 98mm (10.4" x 8.5" x 3.9")
-Standalone weight: 3 kg (6.6 lbs) including battery, RapidPakᵀᴹ clip, NBP cuff, hose, communications cable -and headset:
- Bag size : 401mm x 276mm x 154mm (15.8" x 10.9" x 6.1")
-Weight when packed in bag: 5.9 kg (13 lb) Tempus IC2 plus accessories (excluding power supply, and mains lead/battery charger)
-Thermometer*
-Type: Handheld, Bluetooth-enabled infrared tympanic temperature
- Displayed range:32–43°C (89.6–109.4°F) (target)
- Display resolution: 0.1°C (0.1°F)
- Accuracy: ±0.2ºC (36–39°C) ±0.4°C (<lt/>36 <gt/>39°C) [±0.4ºF (96.8-102.2°F) ±0.5ºF (<lt/>96.8, <gt/>102.2°F)]
- Bluetooth communications range:0–10 m (0–32'9")
12-lead ECG*
- Heart rate range:30–300 bpm
- Input impedance: <gt/>100 MΩ
- Dynamic range: ±5 mV ac
- Accuracy: ±3 %
- DC offset: ±300 mV dc
- Frequency response: 0.05Hz–175Hz ±3 dB⁵
- Sample rate : 500 Hz
-Common mode rejection: 95 dB min, additional filters include mains, muscle, low and high pass and adaptive baseline zeroing
- Protection: Protected against defibrillator charge at 5 kV
-Capnometer - respiration rate:
- Range : 0–149 bpm
- Accuracy: 0–70 bpm: ±1 bpm / 71–121 bpm: ±2 bpm / 122–149 bpm: ±3 bpm
Glucometer:
- Test range: (preset to mg/dL) 10–600 mg/dL (0.6–33.3 mmol/L)
- Reading time: 3 seconds
- Bluetooth communications range: Class 2, range 0–3 m (0–9' 10") in an open field
Pulse oximetry - pulse rate:
- Range : 25-239 bpm
- Accuracy (all ages): No motion ≤3 digits; motion ≤5 digits
Pulse oximetry - SpO₂:
- Range : 1–100 %
- Accuracy (adult/child): No motion/low perfusion: ±2 digits 70–100%; motion: ±3 digits, 70–100%
_Perfusion index range:0.02–20%
Noninvasive blood pressure (NBP):
- Adult/large cuff range : 20–260 mmHg
- Adult/large cuff accuracy: ±3 mmHg or ±2% (whichever is greater)
- Child cuff range: 20–260 mmHg
- Child cuff accuracy: ±3 mmHg or ±2% (whichever is greater)
* Battery and power :
- Charger mains input voltage: 100–240 V
- Charger frequency: 50–60 Hz & 400 Hz
- Charger input currency: 0.9 A max at approximately 100 V
-Charger output voltage:8.4 V dc
- Charger output :current: <lt/> 2.73 A
_Charger weight:0.25 kg nominal
_Charger dimensions:107mm x 67mm x 36.5mm (4.2" x 2.6" x 1.4")
_Power supply, external⁴; PSU-rated 100–250 V, 50 Hz, 115 V 400 Hz
-Battery type: Single, external, user-replaceable, lithium-ion
_Battery strength indicator:Integral
-Battery life: Min 11 hours¹ [default display brightness, SpO₂/EtCO₂ (25% of the time) and NBP every 15 minutes]
- Battery capacity: Nominal 7.4 V, 10.2 Ah
- Battery shelf life: Approximately 7 hours remaining after 1 year storage²
- Battery charge time from empty : 6³ hours
Integral digital camera:
- Annotations : Images can be annotated with text, colors, shapes which can be sent to Tempus IC2 by the hospital